Oxygen analyser on a PSA oxygen plant showing purity dropping from 93% to 88%

Why a ‘93%’ PSA Oxygen Plant Delivers 88%: A Purity-Drop Diagnosis Guide

Introduction

A PSA oxygen generator is commissioned at 93% and runs well for months. Then the purity display starts to read 91%, then 89%, then 88%. For a hospital, 88% is not a small shortfall. It is outside the specification for Oxygen 93.

Purity rarely falls for a mysterious reason. Absstem Shield maintains PSA plants across India, including plants built by other manufacturers (1), and a purity drop almost always traces back to one of five places: the analyser, the demand, the feed air, the valves or the sieve. This guide gives the order in which to check them, the patterns that point to each one, and a real case from the Shield team.

For how the towers and molecular sieves work inside, see our guide to PSA towers (2).

Why 88% Matters

The US Pharmacopeia defines Oxygen 93 Percent as oxygen made from air by the molecular sieve process, containing not less than 90.0% and not more than 96.0% oxygen by volume, the rest mostly argon and nitrogen (3). The WHO International Pharmacopoeia uses the same 90–96% range, often written as 93 ± 3% (4). For more on purity standards in hospitals, see our posts on medical oxygen (5) and oxygen purity in healthcare (6).

So a plant at 88% has left the specification, even though it is still producing oxygen-rich gas. Commissioning guidance for PSA plants requires purity to stay above 90% at all times, checked with a handheld analyser at the tank outlet and at bedside outlets (7).

There is also a ceiling. Air holds about 20.95% oxygen and 0.93% argon, and zeolite does not separate argon from oxygen. The highest purity a PSA oxygen plant can reach is therefore about 95.7%..

Step 1: Is the Reading Real?

Check the analyser before touching the plant. A drifting sensor sends teams chasing problems that do not exist.

  • Check against a second analyser. The Global Fund recommends recording oxygen purity daily with at least two different analysers (8).
  • Check calibration. Calibrate as the analyser maker instructs, and check the reading on ambient air (about 20.9%).
  • Check the sample line. A leaking or wet sample line, or a sample point downstream of a leak, can read low.

If two analysers agree on 88%, the problem is in the plant.

Step 2: Is the Plant Being Overdrawn?

This is the most common cause, and the easiest to fix. A PSA plant trades purity for flow: draw more gas than it was designed for and purity falls. A Global Fund briefing gives the example of a 20 Nm³/h plant at 93 ± 3% that delivers only 18 Nm³/h when set for 95 ± 1% (4).

Signs of overdraw:

  • Purity falls at busy times (ward rounds, cylinder filling, a new production line) and recovers at night.
  • Buffer tank pressure drops below its normal band (9).
  • The product flow meter reads above the rated flow.
  • A new branch, extra cylinder-filling ramp or leak has been added since commissioning.

Fix: find and close leaks, set a flow-control valve at the rated flow, or add capacity. Do not raise the cycle settings to chase flow.

Step 3: Is the Feed Air Right?

The generator can only separate the air it is given. Check four things:

Check What goes wrong Effect on purity
Compressor pressure and free air delivery Worn compressor, slipping belt, clogged intake filter, unloading early Less air per cycle, lower adsorption pressure, purity falls
Air temperature at the generator inlet Aftercooler fouled, hot plant room, dryer bypassed Adsorption efficiency falls as temperature rises (10)
Dew point Refrigerated dryer fault, failed auto drain, liquid water carry-over Water is the critical contaminant for zeolite and reduces its nitrogen capacity (11)
Oil and dust Separator failure, filter elements past their life Coats the sieve, often permanently (2)

Table 1: Feed-air checks for a PSA oxygen purity drop
Dust deserves special mention in Indian conditions. A root-cause study of PSA plants found that dust speeds up the wear of moving parts, causes overheating and restricts airflow (12).

Step 4: Are the Valves and Cycle Working?

Each tower pressurises, produces, equalises and vents many times an hour. If any step is short or leaking, nitrogen breaks through into the product.

  • Uneven purity between towers. Watch the analyser over several cycles. If purity dips every time one particular tower is producing, suspect that tower’s valves or its part of the bed.
  • Leaking or sticking valves. Listen for valves that hiss when closed and check pilot air on pneumatic valves.
  • Clogged exhaust silencer. A blocked silencer slows depressurisation, so the tower does not regenerate fully and purity falls.
  • Timer or controller settings changed. Compare the cycle times with the commissioning record.

Step 5: Is the Sieve Damaged?

Only after Steps 1–4 should the sieve be suspected, because sieve work is the most expensive fix. Signs that point to the sieve:

  • Purity has fallen slowly over months and does not recover at low flow.
  • The fall followed a dryer failure, a power cut with wet air in the line, or an oil event.
  • Dust at the outlet filter, or rising pressure drop across the towers.
  • Frequent sieve topping up, which points to bed movement and attrition (2).

Replacing molecular sieve typically costs 30–40% of the price of the generator (13), which is why protecting the sieve with clean, dry air is far cheaper than replacing it.

Quick Diagnosis Table

Pattern you see Most likely cause Go to
Two analysers disagree Analyser drift or sample line Step 1
Purity low at busy times, fine at night Overdraw or a leak Step 2
Purity low on hot afternoons Feed air too hot Step 3
Purity dropped after a dryer fault or power cut Moisture in the sieve Steps 3 and 5
Purity dips every time one tower produces Valve leak or one bed damaged Steps 4 and 5
Purity falling slowly for months at any flow Sieve contamination or attrition Step 5

Table 2: Purity-drop patterns and where to start

Field Case from Absstem Shield

  • Site: 300-bed district hospital in Northeast India
  • Plant: Absstem 500 LPM plant that’s been  running for 6 years
  • What the hospital saw: The purity reading had gone down – from 93% to 88% slowly, within two weeks
  • What we checked first: We checked the analyser first.
  • Root cause: We realized that the Oxygen sensor’s life was over and that the oxygen purity was, in actuality, 93% itself
  • Fix: We replaced the oxygen sensor
  • Result: As soon as we replaced the sensor, the purity reading was back to being correct – in fact it reached 94%

What a Purity Drop Costs

When a hospital plant goes out of specification, the hospital switches to cylinders. A 50 Nm³/h plant supplies about 1,200 Nm³ a day, which is about 171 standard 7 m³ cylinders. At ₹350 a cylinder, that is about ₹60,000 a day in backup oxygen.

We ran this through the Existing Maintenance vs Absstem Shield calculator in Absstem’s Free Proprietary ROI and Savings Calculator: ₹1.6 lakh a year of maintenance bills and 10 days a year out of specification paid as they come, against a Shield CMC at ₹2.5 lakh a year with 2 days out of specification.

Yearly cost for a 50 Nm³/h plant, paying as you go vs Shield CMC

Figure 1: Yearly cost for a 50 Nm³/h plant, paying as you go vs Shield CMC (Absstem’s Free Proprietary ROI and Savings Calculator)

Parameter Pay as you go Shield CMC
Maintenance or CMC per year ₹1,60,000 ₹2,50,000
Days out of specification 10 2
Backup cylinders and other costs ₹6,18,500 ₹1,23,700
Total per year ₹7,78,500 ₹3,73,700
Saving with CMC – ₹4,04,800 a year (₹20,24,000 over 5 years)

Table 3: Cost of lost purity for a 50 Nm³/h plant (Absstem’s Free Proprietary ROI and Savings Calculator; inputs are illustrations)

At ₹8 per kWh, the calculator puts the power for this plant at ₹7.92 per Nm³, or about ₹9,500 a day of normal running. A day out of specification costs over six times that.

How to Stop It Happening Again

  1. Log purity, flow, buffer pressure, dew point and inlet temperature every day.
  2. Check the analyser against a second unit every week.
  3. Change filter elements and service the dryer on schedule (14).
  4. Keep demand within the rated flow, and add capacity before adding new users.
  5. Keep the plant room cool, ventilated and free of dust.
  6. Use a maintenance contract that includes purity checks, not just breakdown calls. A study of PSA plants found 60% overdue for maintenance at the time of assessment (12). For common misconceptions, see our myths-versus-reality post on PSA maintenance (15).

Internal Resources – Next Steps

Consult Us

For PSA oxygen and nitrogen generators sized to your plant, contact us:

References

  1. Absstem Technologies. (2024). Absstem Shield and Maintenance Contracts for Oxygen Generators.
  2. Absstem Technologies. (2026). Inside a PSA Tower: ZMS, CMS, Sieve Layers and Why Towers Fail.
  3. United States Pharmacopeia. Oxygen 93 Percent (USP–NF monograph).
  4. The Global Fund and Build Health International. (2024). Briefing Note: PSA Plant Oxygen Purity.
  5. Absstem Technologies. (2024). What Is Medical Oxygen? Purity and Standards.
  6. Absstem Technologies. (2024). Importance of Oxygen Purity in Healthcare Applications.
  7. The Global Fund and Build Health International. (2024). PSA Oxygen Plant Commissioning Guidelines.
  8. The Global Fund. PSA Oxygen Plant Sustainability: Overview.
  9. Absstem Technologies. (2024). The Importance of Storage Tank Volume in PSA Systems.
  10. Canadian Patent 1,069,062. Air Separation by Adsorption.
  11. US Patent 7,717,981. Performance Stability in Shallow Beds in Pressure Swing Adsorption Systems.
  12. PATH. Root Cause Analysis of PSA Plant Disrepair.
  13. Absstem Technologies. (2024). How to Choose a PSA Oxygen or Nitrogen Generator.
  14. Absstem Technologies. (2025). PSA Oxygen Plant Maintenance: Schedule, Checks and Repairs.
  15. Absstem Technologies. (2025). PSA Oxygen Plant Repair and Maintenance: Myths vs Reality.

Frequently Asked Questions

Why is my PSA oxygen plant showing low purity?

Check in this order: the analyser, overdraw or leaks, feed air (pressure, temperature, dew point, oil), valves and cycle, and finally the molecular sieve.

Is 88% oxygen acceptable for medical use?

No. Oxygen 93 must contain 90–96% oxygen under USP and the WHO International Pharmacopoeia.

What is the maximum purity of a PSA oxygen plant?

About 95.7%, because argon passes through with the oxygen.

Does hot weather reduce PSA oxygen purity?

es. Zeolite separates less effectively at higher temperature, so hot feed air lowers purity or flow.

Can damaged molecular sieve be repaired?

Mild moisture damage can sometimes be recovered under controlled conditions. Oil contamination is usually permanent and the sieve has to be replaced.

ROI Calculator
×

Scroll to Top